Features
Industry-leading features to maximize productivity


Direct Electron Detection with the DDD technology

  • No resolution degradation from the point spread of a scintillator
  • No fiber optic or other transfer optics to introduce distortions or to scatter electrons
  • Very high signal-to-noise ratio (SNR) for primary electrons (>30:1) enables single electron counting

Small Pixel Size (6 μm)

  • High resolution at low microscope magnification enables film-like resolution and field of view with fast mosaic operation


DE12 DQE Curve

DE12 DQE Curve


Fast Continuous Readout (25 ms per 12 Mega-pixel Frame, >500 fps sub-arrays)

  • Specimen drift correction and sample integrity monitoring for dose sensitive specimens
  • Electron counting mode operation

Drift correction of gold particle clusters

Drift corrected cryo-image using gold particles, 25 80 msec images captured in a 2 second burst.


Electron Counting

  • Single electron SNR>30:1 ensures high detection efficiency
  • Fast readout combined with low electron beam dose condition reduces double counting
  • Improved MTF performance


Single electron events recorded on the sensor

Single electron events recorded on the DE-12 sensor